机译:便携式鼓室计MicroTymp的观察者间可重复性。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

OBJECTIVE: To determine the interobserver reliability of tympanograms obtained with the MicroTymp, a portable tympanometer. SETTING: Family medicine teaching unit in a tertiary care hospital. PATIENTS: Thirty-three patients who presented to the ear, nose and throat clinic in August 1990 for an ear problem. INTERVENTION: Three residents in family medicine independently attempted to record with the MicroTymp one tympanogram for the 66 ears. We excluded the results for seven ears for which tympanograms could not be obtained. MAIN OUTCOME MEASURE: Using objective criteria, two family physicians and two residents in family medicine independently classified the 177 tympanograms into five categories (normal, possible effusion, possible perforation, possible tympano-ossicular dysfunction and unclassifiable). Reliability was estimated by means of the kappa (kappa) coefficient on 161 tympanograms from 59 ears for which the interpretation of the three tympanograms agreed. MAIN RESULTS: The interpretation of the three tympanograms agreed for 34 of the 59 ears (0.58) (kappa = 0.52, 95% confidence limits 0.45 and 0.59). There was no significant difference in interobserver reliability between pairs of observers or between symptomatic and asymptomatic ears. CONCLUSIONS: The interobserver reliability of the MicroTymp is moderate. The tympanograms obtained with the instrument should be interpreted in the context of the clinical findings.
机译:目的:确定使用便携式鼓室仪MicroTymp获得的鼓室图在观察者之间的可靠性。地点:三级医院的家庭医学教学单位。患者:1990年8月因耳部疾病而就诊于耳鼻喉科的33例患者。干预:三名家庭医学住院医师独立尝试用MicroTymp记录66只耳朵的一张鼓室图。我们排除了无法获得鼓室图的七只耳朵的结果。主要观察指标:使用客观标准,两名家庭医生和两名家庭医学住院医师分别将177张鼓室图分为五类(正常,可能的积液,可能的穿孔,可能的鼓膜听觉障碍和无法分类)。可靠性是通过对59只耳朵的161个鼓室图上的kappa(kappa)系数进行估算的,这3个鼓室图的解释是一致的。主要结果:59个耳朵中有34个(0.58)的三个鼓室图的解释一致(kappa = 0.52,95%置信限度0.45和0.59)。观察者对之间或有症状和无症状的耳朵之间的观察者间可靠性没有显着差异。结论:MicroTymp的观察者间的可靠性是中等的。使用仪器获得的鼓室图应在临床发现的背景下进行解释。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号