首页> 中文期刊> 《光谱学与光谱分析》 >双铂网雾化和超声雾化进样系统对ICP-AES分析性能的影响

双铂网雾化和超声雾化进样系统对ICP-AES分析性能的影响

         

摘要

Hildebrand grid nebulizer is a kind of improved Babington nebulizer, which can nebulize solutions with high total dissolved solids. And the ultrasonic nebulizer(USN) possesses advantage of high nebulization efficiency and fine droplets. In the present paper, the detection limits, matrix effects, ICP robustness and memory effects of Hildebrand grid and ultrasonic nebulizers for ICF-AES were studied The results show that the detection limits using USN are improved by a factor of 6~23 in comparison to Hildebrand grid nebulizer for Cu, Pb, Zn, Cr, Cd and Ni. With the USN the matrix effects were heavier, and the degree of intensity enhancement and lowering depends on the element line, the composition and concentrations of matrices. Moreover, matrix effects induced by Ca and Mg are more significant than those caused by Na and Mg, and intensities of ionic lines are affected more easily than those of atomic lines. At the same time, with the USN ICP has less robustness. In addition, memory effect of the USN is also heavier than that of Hildebrand grid nebulizer.%从元素检出限、基体效应、ICP稳健性及记忆效应方面实验研究了双铂网雾化与超声雾化两种进样方式对ICP-AES分析性能的影响.结果表明,在超声雾化进样条件下,Cu,Pb,Zn,Cr,Cd,Ni的检出限较双铂网雾化的检出限降低了6~23倍;在基体效应方面,超声雾化进样时基体对元素原子及离子谱线强度的影响均较双铂网雾化系统明显,元素谱线增强/减弱的幅度受元素谱线、基体组成和浓度的影响;Ca和Mg基体对谱线强度的影响强于K和Na基体;离子谱线受基体的影响明显强于原子谱线.超声雾化系统进样时ICP的稳健性要劣于双铂网雾化系统,而且超声雾化系统的记忆效应较双铂网雾化器严重.