Wear Properties of a Pyrolytic Carbon Intel-vertebral Disc Replacement

机译:热解碳-英特尔椎间盘置换术的磨损特性

获取原文
获取原文并翻译 | 示例

摘要

No significant differences or visible pyrocarbon particles were found in evaluations of all six samples from 0 cycles to 10 million cycles of wear simulation. CMM scans confirmed no significant changes in surface morphology following 10 millions cycles. Thus, there is no detectable wear of the Regain™ lumbar nuclear disc replacement after 10 million cycles of articulation in a worst-case clinical scenario. Furthermore, this study establishes a unique process for the wear testing, isolation, and evaluation of pyrolytic carbon implants on bone.
机译:在从0次循环到1000万次磨损模拟的所有六个样品的评估中,没有发现显着差异或可见的热碳颗粒。 CMM扫描证实,经过1000万次循环后,表面形态无明显变化。因此,在最坏的临床情况下,经过1000万次关节运动后,Regain™腰椎间盘置换术没有可检测到的磨损。此外,这项研究为骨上热解碳植入物的磨损测试,隔离和评估建立了独特的过程。