首页> 外文会议>Precision Electromagnetic Measurements (CPEM), 2012 Conference on >Investigating the use of multimeters to measure high value capacitors at low frequencies against the quantized Hall resistance
【24h】

Investigating the use of multimeters to measure high value capacitors at low frequencies against the quantized Hall resistance

机译:研究使用万用表在低频下针对量化的霍尔电阻测量高值电容器

获取原文
获取原文并翻译 | 示例

摘要

Possibilities and limitations of using measurement system based on two sampling multimeters (DVM) for the ratio measurement of two impedances was investigated in MIKES. Calibration of a 1 µF capacitor from the quantized Hall resistance (QHR) was made at low frequencies. A frequency dependence of ceramic 1 µF capacitor in frequency range 12 Hz – 1600 Hz was measured against different reference resistors at low voltage level with the ratios of 1∶1 to 1∶100. A 1 µF capacitor can be measured at low frequency directly against the QHR by using precision DVM with proper estimation of applied corrections.
机译:在MIKES中研究了使用基于两个采样万用表(DVM)的测量系统来测量两个阻抗的比率的可能性和局限性。根据低频霍尔电阻(QHR)对1 µF电容器进行校准。在低电压电平下,以1∶1至1∶100的比率,针对不同的参考电阻,测量了12 Hz – 1600 Hz频率范围内陶瓷1 µF电容器的频率依赖性。通过使用精密DVM并正确估计应用的校正值,可以直接在低频下针对QHR测量1 µF电容器。