首页> 外文会议>Physic of Medical Imaging pt.2; Progress in Biomedical Optics and Imaging; vol.6 no.22 >Evaluation of the Mechanical Stability of a Mega-Voltage Imaging System using a New Flat Panel Positioner
【24h】

Evaluation of the Mechanical Stability of a Mega-Voltage Imaging System using a New Flat Panel Positioner

机译:使用新型平板定位器评估兆伏成像系统的机械稳定性

获取原文
获取原文并翻译 | 示例

摘要

Mega-Voltage systems are used in radiation oncology both for external radiation delivery and patient positioning prior to treatment. A pair of portal images compared with digitally reconstructed radiographs is currently the gold standard for positioning but new developments have made possible the use of Mega-Voltage Cone Beam CT for better 3D setup. The non-ideal imaging geometry of the treatment unit has a direct impact on both methods. It led to the use of a reticule attachment as reference for the scale and the isocenter position on the portal images. The reticule has limited precision and occasionally super-imposes anatomical information. As for Cone Beam, the image quality crucially depends on the knowledge of the scan geometry during the acquisition. The repro-ducibility of the detector position at each angle will affect the image reconstruction and determine how frequently geometrical calibration must be performed. The objectives of this study are to measure the flex of the detector and evaluate its reproducibility. A RID 1640 Perkin Elmer a-Si Flat Panel is installed on a Siemens Primus linear accelerator with a positioner similar the the one used in the Oncor product. Three original methods are used to investigate the behavior in space and time of the imaging system. A reticule and a Plumb Bob tip are placed along the line formed by the isocenter and the source. Their positions projected on the flat panel for different gantry positions are used to calculate the mechanical flex. Projection matrices obtained in a geometrical Cone Beam calibration are also used to quantify the flat panel sagging. Six full sets of data were acquired over a period of 5 months and recorded overall mechanical flexes of 1 and 3 mm for the transversal and longitudinal directions respectively. The absolute magnitude of the flat panel displacement varies slightly with the method used but the discrepancy stays within the laser precision used for alignment. The small standard deviations of the flat panel displacement ( < 1 mm) suggest great stability over time and permits the clinical implementation of patient positioning without the reticule. More experiments on the positioner with the complete set of projection matrices need to be performed to characterize the long-term behavior of the system and to determinate how frequently the Cone Beam calibration needs to be done to conserve image quality. Future work will develop a daily QA protocol to detect possible collisions that would bring the Cone Beam imaging system out of geometrical calibration.
机译:Mega-Voltage系统在放射肿瘤学中用于外部放射输送和患者在治疗前的定位。目前,与数字重建的X射线照片相比,一对门户图像是定位的黄金标准,但新的发展已使Mega-Voltage锥束CT可以用于更好的3D设置。处理单元的非理想成像几何结构直接影响这两种方法。它导致使用网版附件作为比例和门图像上等角点位置的参考。十字线的精度有限,有时会叠加解剖信息。对于锥束,图像质量关键取决于采集过程中扫描几何的知识。在每个角度的检测器位置的可再现性将影响图像重建,并确定必须多久执行一次几何校准。这项研究的目的是测量检测器的弯曲度并评估其可重复性。 RID 1640 Perkin Elmer a-Si平板安装在Siemens Primus线性加速器上,其定位器类似于Oncor产品中使用的定位器。三种原始方法用于研究成像系统在空间和时间上的行为。沿着由等角点和源所形成的线放置了一个标线和一个铅锤尖。他们针对不同机架位置投影在平板上的位置用于计算机械挠度。在几何锥束校准中获得的投影矩阵也可用于量化平板下垂。在5个月的时间内获取了六套完整的数据,分别记录了横向和纵向的总机械挠曲分别为1和3 mm。平板位移的绝对大小随所使用的方法而略有变化,但差异仍在用于校准的激光精度范围内。平板位移的小标准偏差(<1 mm)表明随着时间的推移具有很大的稳定性,并且可以在没有网线的情况下对患者进行临床定位。需要对具有完整投影矩阵集的定位器进行更多实验,以表征系统的长期行为,并确定需要多久进行一次锥束校准以保存图像质量。未来的工作将开发日常的QA协议,以检测可能导致锥束成像系统脱离几何校准的碰撞。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号