In vivo measurement of blood flow in the retina has been made possible with the advent of Fourier domain optical coherence tomography (OCT). Doppler OCT has seen many advances in recent years in algorithms used for quantifying blood flow. We compare the relative retinal blood flow estimates as measured by the standard phase-resolved (PR) algorithm and the more recent moving-scatterer-sensitive (MSS) algorithm as a function of vessel size. We find that the PR-to-MSS flow ratio significantly decreases with decreasing vessel diameter. We also develop a simulation to approximate the scattering from blood cells in tissue and compare the relative blood flow estimates. The flow ratio measured with simulation closely matches that found in vivo. Our simulation predicts that whereas PR underestimates the flow, MSS overestimates it. Our simulation may help to correct for algorithm bias in in vivo retinal flow estimates.
机译:傅里叶域光学相干断层扫描(OCT)的出现使得体内测量视网膜血流成为可能。近年来,多普勒OCT在用于量化血流的算法中已经取得了许多进步。我们比较了通过标准相分辨(PR)算法和较新的移动散射敏感(MSS)算法测得的相对视网膜血流量估计值与血管大小的关系。我们发现PR与MSS的流量比随着容器直径的减小而显着降低。我们还开发了一种模拟方法,以近似估算组织中血细胞的散射并比较相对的血流估计值。用模拟测量的流量比与体内发现的流量比非常匹配。我们的模拟预测,虽然PR低估了流量,但MSS高估了流量。我们的模拟可能有助于纠正体内视网膜血流估计中的算法偏差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号