首页> 外文会议>Optical coherence tomography and coherence domain optical methods in biomedicine XV >Integrated optical coherence tomography - ultrasound system and miniaturized probes for intravascular imaging
【24h】

Integrated optical coherence tomography - ultrasound system and miniaturized probes for intravascular imaging

机译:集成光学相干断层扫描-超声系统和用于血管内成像的微型探头

获取原文
获取原文并翻译 | 示例

摘要

We report on the development of a multimodal optical coherence tomography (OCT) - ultrasound (US) system and miniaturized OCT-US probe for intravascular imaging. Both OCT optical components and a US transducer were integrated into a single probe, enabling both OCT and US imaging at the same time. A miniaturized OCT-US probe using a single element transducer was designed with a maximum outer diameter of 0.8 mm, which is suitable for in vivo intravascular imaging. The integrated OCT-US imaging system adopted a two-channel data acquisition card to digitize both OCT and US signals. Simultaneous OCT and US data processing and image display were also achieved using our home-developed software. In vitro OCT and US imaging of human aortic tissue was performed using this multimodal imaging system, which demonstrated the feasibility of the OCT-US system in intravascular imaging and its potential in detection of atherosclerotic plaques.
机译:我们报告了多模态光学相干断层扫描(OCT)-超声(US)系统和用于血管内成像的小型OCT-US探针的发展情况。 OCT光学组件和US传感器都集成在一个探头中,可以同时进行OCT和US成像。设计了使用单元件换能器的微型OCT-US探针,最大外径为0.8 mm,适用于体内血管内成像。集成的OCT-US成像系统采用了两通道数据采集卡来数字化OCT和US信号。使用我们自己开发的软件,还可以同时进行OCT和美国数据处理以及图像显示。使用该多模态成像系统对人主动脉组织进行体外OCT和US成像,证明了OCT-US系统在血管内成像中的可行性以及其在检测动脉粥样硬化斑块中的潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号