Integrated Flexible Optitrode for Neural Stimulation and Recording

机译:集成式柔性电极,用于神经刺激和记录

获取原文
获取原文并翻译 | 示例

摘要

We present an integrated flexible probe for deep brain optical stimulation and electrical recording, termed as "Optitrode". This device features an annular light guide consisting of layered transparent polymer and silica, and is fabricated by a unique process we developed: V-groove guided capillary assembly (VGCA), by which a single or multiple microwires can be integrated inside the light guide with high precision. The Optitrode design promises an ultrahigh length-to-diameter ratio > 500 (5 cm long, < 100 μm diameter), enabling minimally-invasive deep brain applications. Furthermore, the highly flexible nature of Optitrode could address a current barrier in rigid brain implants with limited life-time by conforming to the brain micromotion during long-term operation.
机译:我们提出了一种用于深部大脑光学刺激和电记录的集成柔性探头,称为“ Optitrode”。该装置具有由多层透明聚合物和二氧化硅组成的环形光导,并通过我们开发的独特工艺制造而成:V槽导引的毛细管组件(VGCA),通过该器件,单根或多根微线可以集成在光导内部。高精准度。 Optitrode设计保证了超高的长径比> 500(长5 cm,直径<100μm),可实现微创性深部大脑应用。此外,Optitrode的高度灵活特性可以通过在长期手术中顺应大脑的微动,从而在有限的使用寿命内解决刚性脑植入物中的电流障碍。