首页> 外文会议>IEEE International Conference on Biomedical Robotics and Biomechatronics >Design and control of a thermal device for bone cement injection
【24h】

Design and control of a thermal device for bone cement injection

机译:骨水泥注射热装置的设计与控制

获取原文

摘要

This paper presents an original robotized system that reduces potential bone cement leakage during vertebroplasty procedures while providing a longer injection time to radiologists. The system design meets the requirements of vertebroplasty procedures in terms of injection speed, consumables and medical environment. Our contribution relies on the design, dimensioning and control of a device regulating the injected cement temperature, that is one of the main factors affecting the evolution of the cement viscosity. To control the system, thermal exchanges are first modeled from the physical equations of heat and mass transfer. Then, a simple control method is presented and experimentally assessed to confirm the ability to influence the flowing cement temperature at the center of the stream within an acceptable time.
机译:本文介绍了一个原始的机器化系统,可在椎体成形术期间减少潜在的骨水泥泄漏,同时为放射科医师提供更长的注射时间。在注射速度,耗材和医疗环境方面,系统设计符合椎体成形术手续的要求。我们的贡献依赖于调节注入水泥温度的装置的设计,尺寸和控制,这是影响水泥粘度演变的主要因素之一。为了控制系统,首先从热量和传质的物理方程建模热交换。然后,提出和实验评估了简单的控制方法,以确认在可接受的时间内影响流的中心处的流动水泥温度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号