首页> 外文会议>2016 6th IEEE International Conference on Biomedical Robotics and Biomechatronics >Design and control of a thermal device for bone cement injection
【24h】

Design and control of a thermal device for bone cement injection

机译:骨水泥注射加热装置的设计与控制

获取原文
获取原文并翻译 | 示例

摘要

This paper presents an original robotized system that reduces potential bone cement leakage during vertebroplasty procedures while providing a longer injection time to radiologists. The system design meets the requirements of vertebroplasty procedures in terms of injection speed, consumables and medical environment. Our contribution relies on the design, dimensioning and control of a device regulating the injected cement temperature, that is one of the main factors affecting the evolution of the cement viscosity. To control the system, thermal exchanges are first modeled from the physical equations of heat and mass transfer. Then, a simple control method is presented and experimentally assessed to confirm the ability to influence the flowing cement temperature at the center of the stream within an acceptable time.
机译:本文介绍了一种原始的机器人系统,该系统可减少椎骨成形术过程中潜在的骨水泥渗漏,同时为放射科医生提供更长的注射时间。该系统的设计在注射速度,耗材和医疗环境方面均满足椎骨成形术的要求。我们的贡献取决于调节注入水泥温度的设备的设计,尺寸和控制,这是影响水泥粘度变化的主要因素之一。为了控制系统,首先根据热和质量传递的物理方程对热交换进行建模。然后,提出了一种简单的控制方法,并进行了实验评估,以确认在可接受的时间内影响流中心水泥流动温度的能力。