Optimization of ~(99m)Tc-sestamibi/~(123)I subtraction SPECT/CT protocol for parathyroid scintigraphy

机译:优化〜(99m)TC-Sestamibi /〜(123)I减法SPECT / CT协议用于甲状旁腺闪烁扫描

获取原文

摘要

The purpose of this study was to optimize effective, but technically challenging ~(99m)Tc-sestamibi/~(123)I subtraction SPECT/CT protocol for parathyroid scintigraphy. An anthropomorphic parathyroid phantom was set up using a small sphere, a thyroid phantom and a thorax phantom with clinical range of activities of ~(123)I and~(99m)Tc. SPECT/CT acquisitions were performed using three collimators (Low Energy High Resolution (LEHR), Low Energy Ultra High Resolution (LEUHR) and Medium Energy Low Penetration (MELP)) and two energy window settings. Images were reconstructed with a combination of four different numbers of iterations and with or without scatter correction. Images were subjected to visual and quantitative evaluation. The effect of collimator, energy window selection and reconstruction parameters had a significant effect on visual appearance and adenoma contrast in parathyroid ~(99m)Tc-sestamibi/~(123)I subtraction SPECT/CT. Symmetrical energy windows and ultra-high resolution collimator yielded best results with some improvement with scatter correction.
机译:本研究的目的是优化有效,但技术上挑战〜(99米)TC-Sestamibi /〜(123)I减法SPECT / CT协议,用于甲状旁腺闪烁扫描。使用小球体,甲状腺模型和胸腔幻影建立拟人甲状旁腺幻像,临床范围为〜(123)I和〜(99m)Tc。使用三个准直器(低能量高分辨率(LEONR),低能量超高分辨率(LEUHR)和中型能量低渗透(MELP))和两个能量窗口设置进行SPECT / CT采集。以四种不同数量的迭代和有或没有分散校正的组合重建图像。对图像进行视觉和定量评估。准直器,能量窗口选择和重建参数的效果对甲状旁腺〜(99m)TC-Sestamibi /〜(123)I减法Spect / Ct中的视觉外观和腺瘤对比产生显着影响。对称能量窗口和超高分辨率准直器产生了最佳结果,随着散点校正的一些改进。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号