首页> 外文会议>ICAMR 2012;International Conference on Advanced Material Research >The Effect of Denser Material than Air in front of Clinical Radiotherapy X-Ray Beam
【24h】

The Effect of Denser Material than Air in front of Clinical Radiotherapy X-Ray Beam

机译:密集材料比临床放疗前X射线束前的空气的影响

获取原文

摘要

scatter photon produced during radiationtherapy with high energy photons is the main source of unwanted out-of-field and superficial received doses of patients. Surface buildup dose is dependent on electron contamination primarily from the unblocked view of the flattening filter and secondarily from air and collimation systems. We performed a comprehensive set of surface and buildup dose measurements with a thin window parallel-plate (PPC-40) chamber to examine effects of attenuating media in front of 6 MV X ray. To evaluate the impact of beam segmentation on buildup dose, measurements were performed with 10 × 10 cm2 fields, Measurements were performed in Solid Water using parallel plate chambers and diode for a 6 MV X-ray beam.
机译:在具有高能量光子的放射治疗过程中产生的散射光子是不希望的外场和表面接受剂量的患者的主要来源。 表面堆积剂量依赖于电子污染,主要来自扁平滤波器的未封闭视图,并二次来自空气和准直系统。 我们通过薄窗平行板(PPC-40)室进行了一组综合的表面和积聚剂量测量,以检查6 MV X射线前面的衰减介质的效果。 为了评估光束分割对积聚剂量的影响,用10×10cm 2场进行测量,使用平行板室和二极管在固体水中进行测量,用于6mV X射线束。