Validation of the Empatica E4 Wristband

机译:验证EMPatica E4腕带

获取原文

摘要

A preliminary study into the signal quality of the Empatica's E4 portable photoplethysmogram device as validation for further research on using the device to detect the heart arrhythmia atrial fibrillation. In order to pursue this research, it was necessary to verify the quality of the data produced by the E4 device against a device currently used by clinicians to detect atrial fibrillation. This was done by having healthy volunteers wear both the Empatica E4 and the clinician standard device while following standard clinician procedure for atrial fibrillation diagnosis. The device was compared qualitatively against General Electric's SEER Light Extend Recorder holter portable electrocardiogram. The analysis was done by non-experts at varying academic levels. The primary reviewer's results showed similar data quality between both devices 85% of the time and the holter performed better 5% of the time.
机译:EMPatica的E4便携式光电电瓶仪器信号质量的初步研究作为进一步研究使用该装置检测心脏心律失常颤动的验证。为了追求这项研究,有必要验证E4器件对临床医生目前用于检测心房颤动的设备产生的数据的质量。这是通过让健康的志愿者佩戴EMPatica E4和临床医生标准装置,同时进行心房颤动诊断的标准临床手术。定性比较该装置,用于通用电气的Seer光延伸记录器Holter便携式心电图。分析由非专家在不同的学术水平上完成。主要审阅者的结果显示,两种设备之间的数据质量相似85%的时间,并且孔洞的时间更好地执行了5%的时间。