【24h】

CT-guided lung biopsy: Assessment of fiducials registration accuracy

机译:CT引导肺活检:基于基准注册准确性的评估

获取原文

摘要

CT-guided interventions have been widely used in lung biopsy procedures to minimize invasion and to improve biopsy accuracy. The basic mathematical statement of the point-based registration problem remains the same in most applications, whereas there are variations on the size and shape of the fiducial markers. In this paper, we measured registration error FLEp and TRE respectively using two types of fiducial markers, ectrocardiograph electrode and steel bead. The TRE using ECG electrodes is lower than using steel beads. These experimental findings of registration accuracy indicate that ECG electrodes as fiducial markers can be accepted in the lung biopsy.
机译:CT引导的干预已被广泛用于肺活检程序,以最大限度地减少侵袭并提高活检精度。基于点的注册问题的基本数学陈述在大多数应用中保持不变,而基于基准标记的大小和形状存在变化。在本文中,我们分别使用两种类型的基准标记,Ectrocardograph电极和钢珠测量登记误差Flep和Tre。使用ECG电极的TRE低于使用钢珠。这些登记精度的实验结果表明ECG电极可以在肺活组织检查中接受作为基准标记。