首页> 外文会议>2013 IEEE International Conference on Medical Imaging Physics and Engineering >CT-guided lung biopsy: Assessment of fiducials registration accuracy
【24h】

CT-guided lung biopsy: Assessment of fiducials registration accuracy

机译:CT引导下的肺活检:基准点注册准确性评估

获取原文
获取原文并翻译 | 示例

摘要

CT-guided interventions have been widely used in lung biopsy procedures to minimize invasion and to improve biopsy accuracy. The basic mathematical statement of the point-based registration problem remains the same in most applications, whereas there are variations on the size and shape of the fiducial markers. In this paper, we measured registration error FLEp and TRE respectively using two types of fiducial markers, ectrocardiograph electrode and steel bead. The TRE using ECG electrodes is lower than using steel beads. These experimental findings of registration accuracy indicate that ECG electrodes as fiducial markers can be accepted in the lung biopsy.
机译:CT引导的干预已广泛用于肺活检过程中,以最大程度地减少侵袭并提高活检准确性。在大多数应用中,基于点的配准问题的基本数学陈述仍然相同,而基准标记的大小和形状却有所不同。在本文中,我们分别使用心电图电极和钢珠这两种基准标记来测量套准误差FLEp和TRE。使用ECG电极的TRE低于使用钢珠的TRE。配准准确性的这些实验发现表明,在肺活检中可以接受ECG电极作为基准标记。