摘要

Insulation wear in implanted cardioverter-defibrillator (ICD) leads may result in serious failure modes, including inappropriate or missing therapy. Tests performed on a custom apparatus developed to simulate in-vivo loading showed order of magnitude differences in specific wear rate between materials commonly used in ICD leads. Wear rates were also dependent on load, suggesting that design and application differences may further differentiate performance between materials.
机译:植入的心脏除颤器(ICD)引线中的绝缘磨损可能导致严重的失效模式,包括不适当或缺少治疗。在开发的定制装置上执行的测试,以模拟体内载荷的常识在ICD引线中常用的材料之间的幅度差异顺序。磨损率也取决于负荷,表明设计和应用差异可能进一步区分材料之间的性能。