首页> 外文会议>International Symposium on Current Progress in Mathematics and Sciences >Implementation of Small Field Radiotherapy Dosimetry for Spinal Metastase Case
【24h】

Implementation of Small Field Radiotherapy Dosimetry for Spinal Metastase Case

机译:脊髓转移酶壳体小田放射治疗剂量的实施

获取原文

摘要

The main objective of this study was to know dose profile of small field radiotherapy in the spinal metastase case with source axis distance (SAD) techniques. In addition, we evaluated and compared the dose planning of stereotactic body radiation therapy (SBRT) and conventional techniques to measurements with Exradin A16 and Gafchromic EBT3 film dosimeters. The results showed that film EBT3 had a highest precision and accuracy with the average of the standard deviation of ±1.7 and maximum discrepancy of 2.6 %. In addition, the average value of Full Wave Half Maximum (FWHM) and its largest deviation in small field size of 0.8 x 0.8 cm~2 are 0.82 cm and 16.3 % respectively, while it was found around 2.36 cm and 3 % for the field size of 2.4 x 2.4 cm~2. The comparison between penumbra width and the collimation was around of 37.1 % for the field size of 0.8 x 0.8 cm~2, while it was found of 12.4 % for the field size of 2.4 x 2.4 cm~2.
机译:本研究的主要目的是了解源轴距离(悲伤)技术的脊柱转移酶外壳中的小型场放射剂量的剂量型材。此外,我们评估并比较了立体定向体放射治疗(SBRT)和常规技术的剂量计划,以用Exradin A16和GafChromic EBT3薄膜剂量计测量。结果表明,薄膜EBT3具有最高的精度和精度,平均标准偏差为±1.7,最大差异为2.6%。此外,全波半最大(FWHM)的平均值分别为0.8×0.8cm〜2的小场尺寸的最大偏差分别为0.82厘米和16.3%,而该场发现约2.36厘米和3%。大小为2.4 x 2.4 cm〜2。 Penumbra宽度与准直之间的比较为5.1×0.8cm〜2的场尺寸为约37.1%,而现场尺寸为2.4 x 2.4 cm〜2的距离为12.4%。