The Detector Characteristics for Output Factor Measurement of Small Field Electron Beams

机译:小型现场电子束输出因子测量的检测器特性

获取原文

摘要

Have been performed the output factor measurements of the small square field with dimensions of 8 x 8 cm~2 and smaller, for electron beams with energy of 6, 8, 10, 12, and 15 MeV. The experiments has been performed using the Elekta Synergy Platform Linac, solid water, the Exradin A11 plan-parallel ion chamber, Exradin A16 micro ion chamber, PTW Freiburg T60010M-4 silicon diode, and Gafchromic~? EBT-3 film. The result described that the output factor of all detectors decreased with decreasing field size and energy electron beams. The comparison of output factor to output at reference condition for all detectors has discrepancy maximum in the range of 49.5-87.6 % on field 1x1 cm~2 for 6 MeV, and a minimum value in the range of of 0.49-1.21 % on 8 x 8 cm~2 for 15 MeV so it makes the dose distribution of electron treatment of square field of 1 x 1 cm~2 were not appropriate for patient treatment. In this study, PTW Freiburg T60010M-4 silicon diode and Gafchromic? EBT-3 film demonstrated more compatible for small field electron beam measurement.
机译:已经进行了小平方场的输出系数测量,其尺寸为8×8cm〜2且更小,用于具有6,8,10,12和15mev的能量的电子束。实验已经使用ELEKTA协同平台LINAC,固体水,exRADIN A11平行离子室,EXRADIN A16微离子室,PTW Freiburg T60010M-4硅二极管,以及GafChromic〜? EBT-3薄膜。结果描述了所有探测器的输出系数随着场尺寸和能量电子束的降低而降低。所有探测器的参考条件下输出因子对输出的比较最大值在1×1cm〜2的差异范围内为49.5-87.6%,持续6mev,最小值为8 x×0.49-1.21%的值15 meV 8厘米〜2,因此它使得患者治疗的方形场的电子处理剂量分布不合适。在本研究中,PTW弗赖堡T60010M-4硅二极管和GafChromic? EBT-3薄膜对小型现场电子束测量相兼容。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号