首页> 外文会议>International Nuclear Science, Technology and Engineering Conference >Evaluation of radiation dose in pediatric head CT examination: a phantom study
【24h】

Evaluation of radiation dose in pediatric head CT examination: a phantom study

机译:小儿头CT检查中辐射剂量评价:幻影研究

获取原文

摘要

The aim of this study was to evaluate the radiation dose in pediatric head Computed Tomography examination. It was reported that decreasing tube voltage in CT examination can reduce the dose to patients significantly. A head phantom was scanned with dual-energy CT at 80 kV and 120 kV. The tube current was set using automatic exposure control mode and manual setting. The pitch was adjusted to 1.4, 1.45 and 1.5 while the slice thickness was set at 5 mm. The dose was measured based on CT Dose Index (CTDI). Results from this study have shown that the image noise increases substantially with low tube voltage. The average dose was 2.60 mGy at CT imaging parameters of 80 kV and 10-30 mAs. The dose increases up to 17.19 mGy when the CT tube voltage increases to 120 kV. With the reduction of tube voltage from 120 kV to 80 kV, the radiation dose can be reduced by 12.1 % to 15.1% without degradation of contrast-to-noise ratio.
机译:本研究的目的是评估小儿头脑检查断层扫描检查中的辐射剂量。据报道,CT检查中的管电压降低可以显着降低剂量给患者。用80kV和120kV的双能CT扫描头部幻影。使用自动曝光控制模式和手动设置设定管电流。将沥青调节至1.4,1.45和1.5,同时将切片厚度设定为5mm。基于CT剂量指数(CTDI)测量剂量。本研究的结果表明,图像噪声基本上随着低管电压而增加。 CT成像参数为80 kV和10-30 mas的平均剂量为2.60 mg。当CT管电压增加到120 kV时,剂量增加到17.19 MGO。随着120kV至80kV的管电压的降低,辐射剂量可以减少12.1%至15.1%而不会降低对比度噪声。