A Novel Multi-Dimension Force Platform Calibration Design and Error Analysis

机译:一种新型多维力平台校准设计及误差分析

获取原文
获取外文期刊封面目录资料

摘要

Multi-dimension force platform (MFP) is commonly used in human balance function assessment to measure the ground reaction force and the center of pressure (COP). For improving the precision of measurement and simplification the calibration process, this paper devised a generalized load calibration device which can calibrate six axis force simultaneously. Based on the load shift principle, the true loads in every axis were determined when a generalized load was applied on the platform. After experimental verification on force plate devised by our laboratory, the calibration matrix C was calculated based on least-squares approach. The result showed that the crosstalk was kept below the thresholds of most common, commercial force platform.
机译:多维力平台(MFP)通常用于人平衡函数评估,以测量地面反作用力和压力中心(COP)。为了提高测量和简化的精度校准过程,本文设计了一种广义负载校准装置,可以同时校准六个轴力。基于负载换档原理,当在平台上应用广泛的负载时,确定每个轴的真正负载。在实验室设计的力板上进行实验验证后,基于最小二乘法计算校准矩阵C.结果表明,串扰保持在最常见的商业力平台的阈值以下。