首页> 外文会议>International Workshop on Clinical Image-Based Procedures: From Planning to Intervention >Technical Evaluation of a Third Generation Optical Pose Tracker for Motion Analysis and Image-Guided Surgery
【24h】

Technical Evaluation of a Third Generation Optical Pose Tracker for Motion Analysis and Image-Guided Surgery

机译:第三代光学姿势跟踪器的运动分析和图像引导手术技术评估

获取原文

摘要

Laparoscopic instrument tracking systems are an essential component in image-guided interventions and offer new possibilities to improve and automate objective assessment methods of surgical skills. In this study we present our system design to apply a third generation optical pose tracker (Micron-Tracker) to laparoscopic practice. A technical evaluation of this design is performed in order to analyze its accuracy in computing the laparoscopic instrument tip position. Results show a stable fluctuation error over the entire analyzed workspace. The relative position errors are 1.776±1.675 mm, 1.817±1.762 mm, 1.854±1.740 mm, 2.455±2.164 mm, 2.545±2.496 mm, 2.764±2.342 mm, 2.512±2.493 mm for distances of 50, 100, 150, 200, 250, 300, and 350 mm, respectively. The accumulated distance error increases with the measured distance. The instrument inclination covered by the system is high, from 90 to 7.5 degrees. The system reports a low positional accuracy for the instrument tip.
机译:腹腔镜仪器跟踪系统是图像引导干预中的重要组成部分,提供了改进和自动化外科技能的客观评估方法的新可能性。在这项研究中,我们介绍了我们的系统设计,以将第三代光学姿势跟踪器(微米追踪器)应用于腹腔镜实践。进行该设计的技术评估,以分析其在计算腹腔镜仪器尖端位置时的精度。结果在整个分析的工作空间中显示出稳定的波动误差。相对位置误差为1.776±1.675 mm,1.817±1.762 mm,1.854±1.740 mm,2.455±2.164 mm,2.545±2.496 mm,2.764±2.342 mm,2.512±2.493 mm,距离为50,100,150,200,分别为250,300和350 mm。累计距离误差随测量距离而增加。系统涵盖的仪器倾向高,从90到7.5度。系统报告仪器尖端的低位置精度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号