首页> 外文会议>International Conference on 3D Radiation Dosimetry >Comparison of 2D and 3D Gamma calculations for an IMRT QA phantom
【24h】

Comparison of 2D and 3D Gamma calculations for an IMRT QA phantom

机译:对IMRT QA幻影的2D和3D伽马计算的比较

获取原文

摘要

Gamma index pass rates were evaluated for an anthropomorphic phantom using both 2D and 3D calculations. The phantom was irradiated with the traditional dosimetry insert loaded with radiochromic film and TLD, and then with a 3D dosimetry insert. A comparison with the calculated dose distribution showed that both the PRESAGE dosimeter and the film and TLD system agreed with the plan to within 5% using 2D gamma index criteria. The 3D gamma index showed a slightly higher pass rate than the 2D gamma index at 3%/3mm, and comparable pass rates using more generous constraints. The acceptable number of pixels passing the tighter constraints (3%/3mm) might be dependent upon the choice between a 2D calculation versus a 3D calculation.
机译:使用2D和3D计算评估伽马指数通率的拟人幽灵。用装有放射褐色薄膜和TLD的传统剂量件插入物照射幽灵,然后用3D剂量刀片。与计算剂量分布的比较显示,预设剂量表和薄膜和TLD系统同意使用2D伽马索引标准在5%内同意的计划。 3D伽马指数显示比3%/ 3mm的2D伽马指数略高,并且使用更大量的约束的可比传递速率。通过更严格约束(3%/ 3mm)的可接受数量的像素可能取决于2D计算与3D计算之间的选择。