The development of a new solid target station for a medium energy medical cyclotron for proton bombardment of target material perpendicular to the incident proton beam with greater than 80 μA at 11.5 MeV, with a current density of>200 μA/cm~2. A thermocouple is imbedded into the target body to provide constant temperature monitoring during an irradiation while preserving the shuttle transfer mechanism for the target material. The new solid target increases the effective beam current on target from <50 % to>70 % by reducing the beam loss due to divergence along the beam-line.
机译:用于介质能量医疗回旋加速器的新固体目标站的开发,用于垂直于11.5meV大于80μA的入射质子束的靶材料轰击的质子轰击,电流密度>200μA/ cm〜2。热电偶嵌入目标体内,以在照射期间提供恒定的温度监测,同时保留用于目标材料的梭子转移机构。新的固体目标通过沿着光束线的发散引起的光束损耗来增加目标的有效光束电流从<50%至> 70%。