首页> 外文会议>International Meeting on Chemical Sensors >Chitosan Modified Gold Film Microelectrode for the Determination of Iodide
【24h】

Chitosan Modified Gold Film Microelectrode for the Determination of Iodide

机译:壳聚糖改性金膜微电极用于测定碘化物

获取原文

摘要

A simple electrochemical method for iodide detection using colloid Au/chitosan modified gold film microelectrode was developed. The modified electrode was prepared by coating colloid Au/chitosan membrane on the surface of a gold film microelectrode, whose working area could be easily adjusted as required. The detection was carried out in 0.1 mol·L~(-1) KH_2PO_4 (pH=3) solution, and the obtained Anodic Stripping Voltammetry (ASV) current response increased proportionally with iodide concentration in the range of 3.32×10~(-7) mol·L~(-1)~1.57×10-2 mol·L~(-1) with a correlation coefficient of 0.9996. Compared with the commercial iodide selective electrode, it is more sensitive especially at lower concentrations.
机译:开发了一种简单的使用胶体AU /壳聚糖改性金薄膜微电极的碘化物检测电化学方法。通过在金膜微电极的表面上涂覆胶体AU /壳聚糖膜来制备改性电极,其工作区域可以根据需要容易地调节。检测在0.1Mol·L〜(-1)KH_2PO_4(pH = 3)溶液中进行,并且所得阳极溶出伏安法(ASV)电流响应随碘化物浓度比3.32×10〜(-7 )Mol·L〜(-1)〜1.57×10-2mol·L〜(-1),相关系数为0.9996。与商业碘化物选择性电极相比,特别是在较低浓度下更敏感。