首页> 外文会议>SPIE Conference on Physics of Medical Imaging >Reduction of patient dose in digital mammography: Simulation of low-dose image using computed radiography system and flat panel detector system
【24h】

Reduction of patient dose in digital mammography: Simulation of low-dose image using computed radiography system and flat panel detector system

机译:数字乳房X线照相术中患者剂量的减少:使用计算的射线照相系统和平板探测器系统模拟低剂量图像

获取原文

摘要

To reduce the patients' exposure, several low-dose images are necessary to obtain an image that can be used for diagnosis. However, it is clinically undesirable to expose a patient to multiple exposures in order to obtain an optimal image. The purpose of this study was to simulate a low-dose image from the image generated by a routine-dose. Images of acrylic steps were obtained using multiple doses in digital mammography to generate additional noise. The additional noise was calculated as three different noise sources. This study used the digital mammography system with different detectors. It is computed radiography (CR) system and flat panel detector (FPD) system. This noise was added to take into account the resolution of the X-ray detector using the following filters. The filters were designed based on the presampled modulation transfer function (MTF) and digital MTF containing aliasing. The image simulated using the presampled MTF filter was less similar to an actual low-dose image using the CR system. The image simulated using the digital MTF filter was closer to an actual low-dose image compared to the image simulated using the presampled MTF filter using the CR system. The image simulated using the digital MTF filter of the FPD system was similar to an actual low-dose image. By using the proposed method, we were able to obtain a simulated low-dose image from an image generated by a routine-dose.
机译:为了减少患者的暴露,需要几种低剂量图像以获得可用于诊断的图像。然而,在临床上不希望将患者暴露于多个曝光以获得最佳图像。本研究的目的是从常规剂量产生的图像模拟低剂量图像。在数字乳房X线影像中使用多剂量获得丙烯酸类步骤的图像以产生额外的噪声。附加噪声计算为三种不同的噪声源。本研究使用了不同探测器的数字乳腺X线摄影系统。它是计算的射线照相(CR)系统和平板检测器(FPD)系统。添加了这种噪声以考虑使用以下滤波器的X射线检测器的分辨率。基于预采样的调制传递函数(MTF)和包含别名的数字MTF设计过滤器。使用CR系统的使用预采样MTF滤波器模拟的图像与实际的低剂量图像较小。与使用CR系统使用预采样的MTF滤波器模拟的图像相比,使用数字MTF滤波器模拟的图像更接近实际的低剂量图像。使用FPD系统的数字MTF滤波器模拟的图像类似于实际的低剂量图像。通过使用所提出的方法,我们能够从常规剂量产生的图像获得模拟的低剂量图像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号