A dynamic micro-CT scanner with a stationary mouse bed using a compact carbon nanotube field emission x-ray tube

机译:具有静止鼠标床的动态微型CT扫描仪,使用紧凑型碳纳米管场发射X射线管

获取原文

摘要

In this paper we report the development of a high resolution dynamic micro-computed tomography (CT) scanner with a stationary mouse bed using a compact carbon nanotube (CNT) x-ray tube. The scanner comprises a rotating x-ray tube and detector pair and a stationary and a horizontally positioned small animal bed. The system is optimized for in vivo mouse cardiac imaging. Its performance is evaluated with CT scans of phantoms and free-breathing mice. The modulation transfer function (MTF) at 10% is 5 lp/mm. At single frame acquisition, mouse cardiac micro-CT at 20 msec temporal resolution has been demonstrated by prospectively gating the imaging acquisitions to both respiration and cardiac signals.
机译:在本文中,我们通过使用紧凑型碳纳米管(CNT)X射线管,通过静止小鼠床报告高分辨率动态微计算机断层扫描(CT)扫描仪的开发。扫描仪包括旋转X射线管和检测器对和静止和水平定位的小型动物床。该系统针对体内小鼠心脏成像进行了优化。它的性能通过CT扫描和自由呼吸小鼠评估。调制转移功能(MTF)为10%为5磅/mm。在单帧采集时,通过预期将成像获取与呼吸和心脏信号进行了呼吸和心脏信号来证明在20毫秒的时间分辨率下的小鼠心脏微型CT。