首页> 外文会议>Conference on Medical Imaging: Physics of Medical Imaging >Phase contrast mammography with synchrotron radiation: physical aspects of the clinical trial
【24h】

Phase contrast mammography with synchrotron radiation: physical aspects of the clinical trial

机译:同步辐射相位对比乳房X线照相术:临床试验的物理方面

获取原文
获取外文期刊封面目录资料

摘要

Purpose: The first clinical facility for synchrotron radiation (SR) mammography is now operative at the SYRMEP beamline of ELETTRA, the SR facility in Trieste, Italy. The mammographic facility and the preliminary results of the clinical trial are presented in this contribution. Method and Materials: The distance between the SR source and the patient is about 30 m; the main features of the X-ray beam are: monochromaticity at ~0.2% bandwith in the energy range 8-35 keV, photon flux of about 10~8 ph/(mm~2 s) and dimensions of 21 cm * 3.5 mm at the compressed breast. An innovative dosimetric system allows the on-line dose control during the examination. The images are acquired by scanning the patient, in prone position, in front of the stationary laminar beam; the average scanning time is about 10 s. The detector is a screen film system; it is at ~2 m from the breast in order to fulfil the so-called Phase Contrast (PhC) requirements. The breast thickness and glandularity defines the optimal beam energy for each examination. The patients are enrolled by radiologists, after routine examinations, on the basis of BI-RADS classification, according the research program approved by the local Ethical Committee. Results: This communication concerns the first 9 patients underwent the SR PhC mammography; the images match the quality obtained in previous in vitro studies. With reference to conventional mammography the diagnostic quality of the radiological images is better, without increasing the delivered dose to the patient.
机译:目的:同步辐射(SR)乳房X线照相术的第一临床设施现在在Elettra的Syrmep Beamline of Elettra,SR设施在意大利的里雅斯特。乳房Xmmopare设施和临床试验的初步结果在这一贡献中提出。方法和材料:SR源和患者之间的距离约为30米; X射线梁的主要特征是:在能量范围内为8-35keV,光子通量约为10〜8 pH /(mm〜2 s)的单色度,约为10〜8〜2秒,尺寸为21厘米* 3.5mm压缩的乳房。创新的剂量测量系统允许在检查期间进行在线剂量控制。通过在固定层梁前面扫描患者,在俯卧位上扫描患者的图像;平均扫描时间约为10秒。探测器是屏幕薄膜系统;它距离乳房〜2米,以满足所谓的相对对比(PHC)要求。乳房厚度和腺体定义每个检查的最佳光束能量。根据当地道德委员会批准的研究计划,患者在常规考试后由常规检查进行疗程征集。结果:此通信涉及前9例患者进行SR PHC乳房X线图像与先前的体外研究中获得的质量匹配。参考常规乳房X线照相术,放射性图像的诊断质量更好,而不增加患者的递送剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号