Clinical and Statistical Evaluation of 1st Automatic: A Pilot Study

机译:第一天动临床和统计评估:试验研究

获取原文

摘要

Arterial blood gas analyses provide information about the patient's oxygenation, ventilation, and perfusion. Studies have shown that most errors occur during the pre-analytical phase (the processes prior to the actual analysis of the sample). 1st Automatic is an extra module to the existing blood gas analyzer and a partial automation of the pre-analytical phase which is designed to reduce these errors. The objective was to investigate whether 1st Automatic is compatible to the conventional method. Quantitative measurements of pH, pCO2, pO2, Na+, K+, Ca2+, cGlu, cLac and tHb were performed on 21 paired whole blood specimens collected into the standard blood gas syringes and safePICO syringes, both designed for the ABL800 FLEX. During the collection the pre-analytical errors were minimised. No clinically significant differences were observed in the nine blood gas parameters. The differences were statistically insignificant, with the exception of cNa+, cCa2+ and tHb. There was a good consistency between the results using the new automated procedure and the conventional method. It was not possible to show any clinically significant difference between the two procedures.
机译:动脉血气分析提供有关患者氧合,通风和灌注的信息。研究表明,在预分析相期间发生大多数误差(在样品的实际分析之前的过程)。第一天动是现有血气分析仪的额外模块和预分析阶段的部分自动化,旨在减少这些误差。目的是研究第1天动是否与传统方法兼容。在将21个成对的全血样标本上进行,对PH,PCO2,PO2,Na +,K +,Ca2 +,CgLU,ClAC和THB进行定量测量,用于将其集中成标准血液气体注射器和SafePico注射器,两者都针对ABL800 Flex设计。在收集期间,预分析误差是最小化的。在九个血气参数中没有观察到临床上显着的差异。差异是统计学上微不足道的,除CNA +,CCA2 +和THB外。使用新的自动化程序和传统方法的结果之间存在良好的一致性。在两种程序之间无法显示任何临床上有显着差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号