首页> 外文会议>International Conference on Synchrotron Radiation Instrumentation >Properties of Wide-dose-range GafChromic Films for Synchrotron Radiation Facility
【24h】

Properties of Wide-dose-range GafChromic Films for Synchrotron Radiation Facility

机译:Synchrotron辐射设施宽剂量范围Gafchromic薄膜的性质

获取原文

摘要

GafChromic films have been used at SPring-8 to detect the intensively irradiated parts and protect them from damage by being covered with shield or moved. To extend the usable dose range more widely, a new type of sensitive film EBT was investigated for the introduction. Calibration curves were obtained irradiated with 60Co rays and compared with those of other GafChromic films. For the application, these films were set in the white x-ray hutch and the dose distribution was measured. Ratio of doses given by EBT and XT-R indicated the degree of the photon spectrum hardness, which depended on the positions. As a result, dose range from 50 mGy to 300 kGy became available for dose distribution measurements, and a set of films having different energy responses was found to give information of photon spectra.
机译:Grafchromic薄膜已在弹簧-8中使用,以检测强烈照射的部件,并通过屏蔽或移动来保护它们免受损坏。为了更广泛地扩展可用剂量范围,研究了一种新型的敏感薄膜EBT进行引入。用60Co射线照射校准曲线,与其他Gafchromic薄膜的射线相比。对于应用,将这些薄膜设置在白色X射线壳中,测量剂量分布。 EBT和XT-R给出的剂量的比例表示光子谱硬度的程度,这取决于位置。结果,可用于剂量分布测量的50 mgy至300kgy的剂量范围,并且发现一组具有不同能量响应的薄膜来提供光子谱的信息。