Measurement of signal-to-noise ratio for a mixture of coherent and incoherent sources using digital sampling oscilloscope

机译:使用数字采样示波器测量相干和不相干源的混合物的信噪比

获取原文

摘要

Signal-to-noise ratios for a coherent laser source and a mixture of a laser source with a narrow band amplified spontaneous emission source are directly measured using a digital sampling oscilloscope. The measurement for a coherent source is found unreliable due to the low level of shot-noise compared to the thermal noise. On the other hand, the measurement for a mixed source is found in good agreement with theoretical predictions within 1dB error.
机译:使用数字采样示波器直接测量相干激光源的信噪比和具有窄带放大的自发发射源的激光源的混合物。由于与热噪声相比,由于镜头噪声水平低,相干源的测量不可靠。另一方面,与1dB误差内的理论预测良好的协议,发现混合源的测量值。