首页> 外文会议>Physics of medical imaging conference >Monte Carlo study of x-ray cross-talk in a variable resolution x-ray detector
【24h】

Monte Carlo study of x-ray cross-talk in a variable resolution x-ray detector

机译:Monte Carlo在可变分辨率X射线探测器中的X射线交叉谈话

获取原文

摘要

A variable resolution x-ray (VRX) detector provides a great increase in the spatial resolution of a CT scanner. An important factor that limits the spatial resolution of the detector is x-ray cross-talk. A theoretical study of the x-ray crosstalk is presented in this paper. In the study, two types of the x-ray cross-talk were considered: inter-cell and inter-arm cross-talk. Both types of the x-ray cross-talk were simulated, using the Monte Carlo method, as functions of the detector field of view (FOV). The simulation was repeated for lead and tungsten separators between detector cells. The inter-cell x-ray cross-talk was maximum at the 34-36 cm FOV, but it was low at small and the maximum FOVs. The inter-arm x-ray cross-talk was high at small and medium FOVs, but it was greatly reduced when variable width collimators were placed on the front surfaces of the detector. The inter-cell, but not inter-arm, x-ray cross-talk was lower for tungsten than for lead separators. From the results, x-ray cross-talk in a VRX detector can be minimized by imaging all objects between 24 cm and 40 cm in diameter with the 40 cm FOV, using tungsten separators, and placing variable width collimators in front of the detector.
机译:可变分辨率X射线(VRX)检测器提供CT扫描仪的空间分辨率的大大增加。限制探测器的空间分辨率的一个重要因素是X射线串扰。本文提出了对X射线串扰的理论研究。在研究中,考虑了两种类型的X射线串扰:细胞间和臂间交叉谈话。使用Monte Carlo方法模拟了两种类型的X射线串扰,作为探测器视野(FOV)的功能。对检测器单元之间的铅和钨分离器重复模拟。细胞间X射线串扰最大在34-36cm foV,但在小而最大的FOV中很低。臂间X射线交叉口交位于中小型FOV,但是当将可变宽度准直器放置在检测器的前表面上时,它大大降低。细胞间,但不是臂间,X射线串扰比铅分离器更低。从结果,通过使用钨分离器将直径24厘米和40厘米的直径介于24厘米和40厘米的直径之间的所有物体进行成像,可以最小化VRX检测器中的X射线交叉口交,并将可变宽度准直器放置在检测器前面。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号