首页> 外文会议>Medical Imaging 2003: Physics of Medical Imaging >Monte Carlo study of x-ray cross-talk in a variable resolution x-ray detector
【24h】

Monte Carlo study of x-ray cross-talk in a variable resolution x-ray detector

机译:可变分辨率X射线探测器中X射线串扰的蒙特卡洛研究

获取原文

摘要

A variable resolution x-ray (VRX) detector provides a great increase in the spatial resolution of a CT scanner. An important factor that limits the spatial resolution of the detector is x-ray cross-talk. A theoretical study of the x-ray crosstalk is presented in this paper. In the study, two types of the x-ray cross-talk were considered: inter-cell and inter-arm cross-talk. Both types of the x-ray cross-talk were simulated, using the Monte Carlo method, as functions of the detector field of view (FOV). The simulation was repeated for lead and tungsten separators between detector cells. The inter-cell x-ray cross-talk was maximum at the 34-36 cm FOV, but it was low at small and the maximum FOVs. The inter-arm x-ray cross-talk was high at small and medium FOVs, but it was greatly reduced when variable width collimators were placed on the front surfaces of the detector. The inter-cell, but not inter-arm, x-ray cross-talk was lower for tungsten than for lead separators. From the results, x-ray cross-talk in a VRX detector can be minimized by imaging all objects between 24 cm and 40 cm in diameter with the 40 cm FOV, using tungsten separators, and placing variable width collimators in front of the detector.
机译:可变分辨率X射线(VRX)检测器极大地提高了CT扫描仪的空间分辨率。限制探测器空间分辨率的重要因素是X射线串扰。本文介绍了X射线串扰的理论研究。在研究中,考虑了两种类型的X射线串扰:单元间串扰和手臂间串扰。使用蒙特卡洛方法,对两种类型的X射线串扰进行了模拟,以作为探测器视场(FOV)的函数。对检测器单元之间的铅和钨隔板进行了重复模拟。细胞间X射线串扰在34-36 cm FOV处最大,但在较小和最大FOV处较低。臂间X射线串扰在中小FOV时较高,但是当将可变宽度的准直器放置在检测器的前表面时,交叉干扰大大降低了。钨的电池间而不是手臂间的X射线串扰低于铅隔板。根据结果​​,通过使用钨分离器在40 cm FOV上对直径在24 cm至40 cm之间的所有物体进行成像,并使用可变宽度的准直仪放置在检测器的前面,可以最小化VRX检测器中的X射线串扰。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号