首页> 外文会议>Meeting of the American Electrophoresis Society >Glucose control during exercise in type I diabetic patients
【24h】

Glucose control during exercise in type I diabetic patients

机译:I型糖尿病患者运动期间的葡萄糖控制

获取原文

摘要

An automated glucose concentration controller for insulin-dependent diabetics was synthesized. The model predictive control framework was used due to its ability to incorporate the effects of past manipulated variable moves into the future move decisions. Controller performance was evaluated in simulation by subjecting the diabetic patient to bouts of mild (30 percent VO_2~(max)) and moderate (60 percent VO_2~(max)) exercise. The controller was capable of limiting glucose concentration undershoot towithin 3 mg/dL of the setpoint, while open-loop responses demonstrated >= 10 mg/dL undershoot. As desired, the controller displayed aggressive response to more strenuous exercise while remaining relatively insensitive to mild exercise disturbances.
机译:合成了胰岛素依赖性糖尿病患者的自动血糖浓度控制器。由于其将过去操纵变量的效果转化为未来的移动决策,因此使用了模型预测控制框架。通过使糖尿病患者进行轻度(30%VO_2〜(MAX))和中等(60%VO_2〜(MAX))运动,评估控制器性能。控制器能够限制设定点的葡萄糖浓度下冲纤维3mg / dl,而开环响应显示> = 10mg / dl下冲。根据需要,控制器显示对更剧烈的锻炼的积极反应,同时保持对温和运动干扰的相对不敏感。