Imaging in Sitting Position May Reduce Liver Artifact in Myocardium Perfusion Imaging

机译:坐姿成像可以减少心肌灌注成像中的肝术

获取原文

摘要

In this work we evaluated liver artifacts in SPECT myocardium perfusion imaging. Liver artifact occurs due to the liver-heart proximity and is manifested as an apparent increase or decrease in radioactive uptake in myocardium. The main motivation for the work is that there are now SPECT imaging devices on the market that image patients in an upright seated position, as opposed to the conventional supine position. The upright position may aid in removal of the artifact during reconstruction by causing the liver to sag downward, thus increasing liver-heart distance. In this work we investigate whether this change of the liver-heart distance can result in more accurate image reconstructions, and thus lower myocardium defect misdetection rates. To investigate this we use the realistic NCAT phantom to simulate the human torso, and SIMIND simulation code.
机译:在这项工作中,我们评估了SPECT心肌灌注成像中的肝术。由于肝心脏接近,肝脏伪影发生,并且表现为心肌中放射性摄取的表观增加或减少。这项工作的主要动机是,现在市场上的SPECT成像装置是在直立坐姿的患者中,而不是传统的仰卧位。直立位置可以帮助在重建期间通过使肝脏向下凹陷来消除伪像,从而增加肝心距离。在这项工作中,我们研究了肝脏心距的这种变化是否会导致更准确的图像重建,从而降低心肌缺陷误差率。为了调查这一点,我们使用现实的NCAT幻影来模拟人类躯干和Simind仿真代码。