A Bilateral Comparison on Humidity Standards in the Dew/Frost-point Temperature Range from -30 deg C to 20 deg C

机译:露水/霜点温度范围内的湿度标准的双边比较从-30℃至20℃

获取原文

摘要

This is a regional metrology organization (SIM) bilateral comparison of the realizations of the scale of dew/frost point temperature at the participating National Metrology Institutes: NIST (USA) and INMETRO (Brazil). A total of four humidity levels in terms of dew/frost-point temperature are used for the comparison: 20 deg C, 0 deg C, -10 deg C and -30 deg C. At each measured value, the mean and standard deviation of multiple readings of the transfer standard, which is a chilled-mirror type of dew-point hygrometer, were recorded. The results of statistical analysis support the conclusion that the measurement data can be assumed to have the form of normal distribution. Combined standard uncertainty ranged from 0.02 deg C to 0.08 deg C approximately.
机译:这是参与国家计量机构的露水/霜点温度规模的实现的区域计量组织(SIM)的双边比较:NIST(美国)和Inforro(巴西)。在每个测量值下,使用露水/霜点温度的四个湿度水平用于比较:20℃,0℃,-10℃和-30℃。平均值和标准偏差记录了传输标准的多读,这是一种冷却镜像的露点湿度计。统计分析结果支持得出测量数据可以假设具有正态分布形式的测量数据。组合的标准不确定性大约为0.02℃至0.08℃。