The single-beam spectra recorded with two commercially available monostatic FT-IR open-path monitors exhibited nonzero signal intensities in the wave number regions where water vapor and CO$-2$/ totally absorb. Each instrument yielded a measureable spectral response with the transmitting/receiving telescope blocked. This response is attributed to stray light within the FT-IR system. In some cases the stray light contributed up to 17% of the total return intensity measured along a 414-m path length. Although the baseline region of the single-beam spectra can be corrected for stray light, its effect on the spectral absorbance is not uniform over the range of absorbance values recorded in long-path measruements. Thus, the concentration of species measured along the path cannot be easily corrected, which has a deleterious effect on the accuracy of the spectral data.
机译:记录有两个商业上可获得的单体FT-IR开放式路径监测器的单光束光谱在水蒸气和CO $ -2 $ /完全吸收的波数区域中表现出非零信号强度。每个仪器产生可测量的频谱响应,透射/接收望远镜被阻挡。该响应归因于FT-IR系统内的杂散光。在某些情况下,杂散光沿着414米的路径长度测量的总返回强度的高达17%。尽管可以校正单光谱光谱的基线区域对于杂散光,但其对光谱吸光度的影响在长路径测量中记录的吸光度值范围内并不均匀。因此,沿着路径测量的物种的浓度不能容易地校正,这对频谱数据的精度具有有害影响。