首页> 外文会议> >Effects of stray light in FTIR instruments on open-path measurements
【24h】

Effects of stray light in FTIR instruments on open-path measurements

机译:FTIR仪器中杂散光对开放路径测量的影响

获取原文

摘要

Abstract: The single-beam spectra recorded with two commercially available monostatic FT-IR open-path monitors exhibited nonzero signal intensities in the wave number regions where water vapor and CO$-2$/ totally absorb. Each instrument yielded a measureable spectral response with the transmitting/receiving telescope blocked. This response is attributed to stray light within the FT-IR system. In some cases the stray light contributed up to 17% of the total return intensity measured along a 414-m path length. Although the baseline region of the single-beam spectra can be corrected for stray light, its effect on the spectral absorbance is not uniform over the range of absorbance values recorded in long-path measruements. Thus, the concentration of species measured along the path cannot be easily corrected, which has a deleterious effect on the accuracy of the spectral data. !0
机译:摘要:用两个市售的单静态FT-IR开放路径监测器记录的单束光谱在水蒸气和CO $ -2 $ /完全吸收的波数区域表现出非零信号强度。每种仪器在发送/接收望远镜均受阻的情况下产生了可测量的光谱响应。该响应归因于FT-IR系统内的杂散光。在某些情况下,杂散光占沿414-m路径长度测得的总返回强度的17%。尽管可以针对杂散光校正单光束光谱的基线区域,但在长距离测量中记录的吸光度值范围内,其对光谱吸光度的影响并不均匀。因此,沿着该路径测量的物质的浓度不能容易地校正,这对光谱数据的准确性具有有害影响。 !0