首页> 外文会议>International conference on monitorning of toxic chemicals and biomarkers >Development of a long-path Fourier transform infrared instrument for measurements of diffuse VOC emissions from the automobile industry
【24h】

Development of a long-path Fourier transform infrared instrument for measurements of diffuse VOC emissions from the automobile industry

机译:用于从汽车工业漫射VOC排放测量的长路径傅立叶变换红外仪器的开发

获取原文

摘要

This paper describes the construction of a long-path fourier transform instrument based on a BOMEM MB100 spectrometer, as well as the initial application for VOC monitoring at an automobile painting facility. The instrument gave detection limits of about 100 ug/m$+3$/ for a number of aromatic hydrocarbons along a 350 m folded measurement path (single end retroreflector setup, 180 scans, 1 cm$+$MIN@1$/ resolution). The instrument also yielded a detection limit of about 10 ug/m$+3$/ for butyl acetate which promotes this compound to be used as a tracer for VOC emissions originating from the automobile industry. The paper also includes a description of the spectral analysis applied involving single component classical least square fit and correction for water vapor interference.
机译:本文介绍了基于Bomem MB100光谱仪的长路径傅立叶变换仪的构造,以及汽车绘画设施的VOC监测初始应用。仪器给出了沿350米折叠测量路径的多个芳烃的检测限为约100ug / m $ + 3 $ /,符合350米的测量路径(单端回转轮设置,180次扫描,1cm $ + $ min @ 1 $ /分辨率) 。该仪器还产生了约10 ug / m + 3美元/乙酸丁酯的检测限,乙酸丁酯促进该化合物用作源自汽车工业的VOC排放的示踪剂。本文还包括涉及单组分经典最小二乘拟合和水蒸气干扰校正的谱分析的描述。