NATIONAL COMPARISON OF SPRING CONSTANT MEASUREMENTS OF ATOMIC FORCE MICROSCOPE CANTILEVERS

机译:国家力量显微镜悬臂弹簧常数测量的全国性比较

获取原文

摘要

This paper illustrates a comparison of spring constant measurements between Technische Universitat Ilmenau (TU Ilmenau) and Physikalisch-Technische Bundesanstalt (PTB). For the traceable comparison, an atomic force microscope (AFM) cantilever with a nominal stiffness of 40 N/m was chosen as the transfer artefact. The determined spring constants differ less than 1%. The measurement uncertainty was ≤4% (k = 2) and hence conforms to the state of the art of AFM cantilever spring constant measurements. The main common uncertainty contribution of TU Ilmenau and PTB is due to the adjustment of the cantilever towards the calibration force.
机译:本文说明了Technische Universitat Ilmenau(Tu Ilmenau)和Physikalisch-Technische Bundesanstalt(PTB)之间的弹簧常数测量的比较。对于可追踪的比较,选择具有40n / m标称刚度的原子力显微镜(AFM)悬臂作为转移伪像。确定的弹簧常数不同小于1%。测量不确定度≤4%(k = 2),因此符合AFM悬臂弹簧恒定测量的领域的状态。 Tu Ilmenau和PTB的主要常见不确定性贡献是由于悬臂朝向校准力的调整。