首页> 外文会议>DAAAM International Symposium on Intelligent Manufacturing and Automation >Improving the Finite Element Simulation of Wear of Total Hip Prosthesis' Spherical Joint with the Polymeric Component
【24h】

Improving the Finite Element Simulation of Wear of Total Hip Prosthesis' Spherical Joint with the Polymeric Component

机译:用聚合物组分改善总髋关节假体球形球形接头磨损的有限元模拟

获取原文

摘要

The advanced model of wear in the spherical joint of total hip prosthesis comprising an acetabular cup of ultra-high molecular weight polyethylene (UHMWPE) in combination with a metal or ceramic femoral head is developed. The wear model is based on the classical Archard-Lancaster equation in common with all other studies reported in literature. The finite element solution of the contact problem between the cup and the head was employed under the loading and angular motions conditions according to the ISO 14242-1 demands. The polymer wear in terms of cumulative linear and volume wear when the wear factor is chosen to be a function of contact pressure is first evaluated.
机译:开发了含有髋臼超高分子量聚乙烯(UHMWPE)的总髋关节前髋关节磨损的先进模型,与金属或陶瓷股头组合。磨损模型基于古典的Archard-Lancaster方程与文献中报告的所有其他研究共同。根据ISO 14242-1要求,使用杯子和头部之间接触问题的有限元溶液。首先评估当磨损因子是磨损因子作为接触压力的函数时,在累积线性和体积磨损方面的聚合物磨损。