首页> 外文会议>IEEE/CPMT International Electronic Manufacturing Technology Conference >Correlation effects of ATE transfers: An empirical study
【24h】

Correlation effects of ATE transfers: An empirical study

机译:ATE转移的相关效应:实证研究

获取原文

摘要

This study presents an empirical analysis of the effects on correlation of the site-to-site transport of ATEs. Four correlation scenarios are presented with permutations of calibration/non-calibration of the testers. The results show that the least number of uncorrelated tests are encountered when there is calibration at the reference tester site, with or without calibration on the target site. Also, there is no significant shift in the distribution of tests after the physical transport and the calibration of the testers. Any shift encountered in these cases can be attributed to a number of factors, such as device-specific sensitivity, vulnerability of related tester resources, or degradation of the representative units used per device. Overall, this study demonstrates that non-calibration at the target site, provided there is calibration at the reference tester site, is a viable and credible option for a less time-consuming and more cost-efficient correlation procedure, especially in massive transfer of ATEs.
机译:本研究提出了对现场与现场运输的相关影响的实证分析。校准/非校准的排列提供了四种相关方案。结果表明,当参考测试仪现场校准时,在参考测试仪现场校准时,遇到了最少数量的不相关测试,或者在目标站点上校准。此外,物理运输后测试的分布和测试人员的校准没有显着转变。在这些情况下遇到的任何转变都可归因于许多因素,例如设备特定的灵敏度,相关测试仪资源的漏洞,或者每个设备使用的代表单位的劣化。总的来说,本研究表明,在参考测试仪现场存在校准的目标场所的非校准是一种可行且可靠的选择,用于较少耗时和更具成本效益的相关程序,特别是在大规模转移的ate 。