Using gas additives to improve signal intensity and selectivity of a cylindrical FAIMS device

机译:使用气体添加剂来提高圆柱形指数装置的信号强度和选择性

获取原文

摘要

The addition of trace gases to a cylindrical FAIMS has been shown to dramatically affect the CV spectra of the small molecules given in Table 1. Changes in CV spectra are greatest at very small levels (≤0.02%) and explains why contaminants must be tightly controlled when operating a cylindrical FAIMS without the use of gas additives. Especially for small molecules, because of the focusing capabilities unique to the cylindrical FAIMS device, dramatic improvements in the signal intensity are observed This workd focused on using acetontrite; a more detailed analysis of other gas additives will be carried out.