A QconCAT Standard for Calibration of Ion Mobility Mass Spectrometry

机译:用于校准离子迁移率质谱法的QConcat标准

获取原文
获取外文期刊封面目录资料

摘要

The physical principles that determine ion mobility in a linear drift cell are well understood (1); consequently, collision cross sections can be directly calculated from the measured drift time having defined the experimental conditions. Such calculations are not possible for Travelling Wave (T-wave) mobility instruments due to the range of factors that influence ion motion through the mobility cell. Calibration of these instruments is thus required for each set of experimental conditions so that T-wave mobility experiments can be used to calculate analyte ion collision cross-sections. Here we describe the design of a universal mobility calibration standard, QCAL-IM, enabling empirical calculation of peptide collision cross-sections using a T-wave mobility instrument.
机译:确定线性漂移细胞中离子迁移率的物理原理很好地理解(1);因此,可以从测量的漂移时间直接计算碰撞横截面,所述测量的漂移时间限定了实验条件。由于影响离子运动通过迁移率的因素的范围,这种计算是不可能的旅行波(T波)移动仪器。因此,每组实验条件需要这些仪器的校准,因此可以使用T波迁移率实验来计算分析物离子碰撞横截面。在这里,我们描述了使用T波移动性仪器的通用移动校准标准,QCAL-IM的设计,使得能够经验计算肽碰撞横截面。