首页> 外文会议>American Society for Mass Spectrometry Conference on Mass Spectrometry and Allied Topics >Automation of Integrated LC-MS/MS Peak Quality Assurance and Analytical Run Quality Control
【24h】

Automation of Integrated LC-MS/MS Peak Quality Assurance and Analytical Run Quality Control

机译:集成LC-MS / MS峰值质量保证和分析运行质量控制的自动化

获取原文

摘要

Take Home Message: For the Pgp Transporter assay we adopted the philosophy of reviewing only those 5-10percent of the samples in a data set that needed manual attention; This was done by using the two automated process queries in MultiQuant to identify the 90-95percent of the samples that passed quality and acceptance criteria and therefore did not need any manual review; By taking this approach we were able to cut down on the number of data points that were being manually reviewed and eliminate the data review step as the bottleneck in the Pgp Transporter assay; Overall: Data for 90 compounds was reported per week, with a weekly data review time savings of 5-6 hours.
机译:带回家消息:对于PGP运输器测定,我们采用了在需要手动注意的数据集中审查样本的5-10%的哲学;这是通过使用多谜语中的两个自动化过程查询来完成的,以识别通过质量和验收标准的样本的90-95percent,因此不需要任何手动回顾;通过采取这种方法,我们能够减少手动审查的数据点数,并消除数据审查步骤作为PGP运输仪测定中的瓶颈;总体:每周报告90种化合物的数据,每周数据审查时间节省5-6小时。