首页> 外文会议>American Society for Mass Spectrometry Conference on Mass Spectrometry and Allied Topics >Performance Evaluation of Three LC/MS Methods Implemented on an Ion Trap Mass Spectrometer for Drug Testing in Urine
【24h】

Performance Evaluation of Three LC/MS Methods Implemented on an Ion Trap Mass Spectrometer for Drug Testing in Urine

机译:三种LC / MS方法对尿液试验中的三种LC / MS方法的性能评价

获取原文

摘要

All 245 compounds screened with MS~(2)/MS~(3) and MS~(3) methods in spiked urine samples were identified. The probability of non-specific endogenous peak ion selection for spectra acquisition increases as the number of targeted compounds in screening method increases. Identification limits in methods with longer scan events cycle (MS~(2)/MS~(3) and MS~(3) methods) are affected more by collecting spectra for endogenous compounds than methods with shorter scan events cycle (MS~(2) method). Methods collecting MS~(2) and MS~(3) or just MS~(3) spectra provide stronger hit confirmation. They are less affected by interference but may result in higher detection limits for some compounds compared to method using just MS~(2) spectra.
机译:鉴定了所有245个用MS〜(2)/ ms〜(3)和MS〜(3)方法筛选的化合物均鉴定在尖刺尿液样品中。随着筛选方法中的靶向化合​​物的数量的增加,非特异性内源性峰离子选择的概率增加。具有较长扫描事件循环的方法的鉴定限制(MS〜(2)/ ms〜(3)和MS〜(3)方法)通过收集内源化合物的光谱而受到比具有较短扫描事件循环的方法(MS〜(2 ) 方法)。方法采集MS〜(2)和MS〜(3)或仅仅是MS〜(3)光谱,提供更强的击中确认。与使用MS〜(2)光谱的方法相比,它们的干扰影响较小,但可能导致一些化合物的检测限。