Ultra High Resolution Linear Ion Trap Orbitrap Mass Spectrometer (Orbitrap Elite) Facilitates Top Down LC MS/MS and Versatile Peptide Fragmentation Modes

机译:超高分辨率线性离子阱Orbitrap质谱仪(Orbitrap Elite)有助于自上而下的LC MS / MS和多功能肽片段化模式

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Although only a few years old, the combination of a linear ion trap with an Orbitrap analyzer has become one of the standard mass spectrometers to characterize proteins and proteomes. Here we describe a novel version of this instrument family, the Orbitrap Elite, which is improved in three main areas. The ion transfer optics has an ion path that blocks the line of sight to achieve more robust operation. The tandem MS acquisition speed of the dual cell linear ion trap now exceeds 12 Hz. Most importantly, the resolving power of the Orbitrap analyzer has been increased twofold for the same transient length by employing a compact, high-field Orbitrap analyzer that almost doubles the observed frequencies. An enhanced Fourier Transform algorithm—incorporating phase information—further doubles the resolving power to 240,000 at m/z 400 for a 768 ms transient. For top-down experiments, we combine a survey scan with a selected ion monitoring scan of the charge state of the protein to be fragmented and with several HCD microscans. Despite the 120,000 resolving power for SIM and HCD scans, the total cycle time is within several seconds and therefore suitable for liquid chromatography tandem MS. For bottom-up proteomics, we combined survey scans at 240,000 resolving power with data-dependent collision-induced dissociation of the 20 most abundant precursors in a total cycle time of 2.5 s—increasing protein identifications in complex mixtures by about 30%. The speed of the Orbitrap Elite furthermore allows scan modes in which complementary dissociation mechanisms are routinely obtained of all fragmented peptides.
机译:尽管只有几年的历史,但线性离子阱与Orbitrap分析仪的结合已成为表征蛋白质和蛋白质组学的标准质谱仪之一。在这里,我们介绍了Orbitrap Elite仪器系列的新版本,它在三个主要方面进行了改进。离子传输光学器件的离子路径会遮挡视线,从而实现更稳定的操作。现在,双池线性离子阱的串联MS采集速度超过12 Hz。最重要的是,通过使用紧凑的高场Orbitrap分析仪,对于相同的瞬变长度,Orbitrap分析仪的分辨能力提高了两倍,而观察到的频率几乎翻了一番。增强的傅里叶变换算法(结合了相位信息)可在768 ms瞬态条件下将m / z 400的分辨能力进一步翻倍至240,000。对于自上而下的实验,我们将调查扫描与要破碎的蛋白质的电荷状态的选定离子监测扫描以及几种HCD显微扫描结合起来。尽管SIM和HCD扫描具有120,000的分辨能力,但总循环时间仍在数秒之内,因此适用于液相色谱串联MS。对于自下而上的蛋白质组学,我们将240,000分辨率的调查扫描与数据相关的碰撞诱导的20种最丰富前体的离解相结合,总循环时间为2.5 s-将复杂混合物中的蛋白质鉴定提高了约30%。 Orbitrap Elite的速度进一步允许扫描模式,在该模式下,常规获得所有片段化肽的互补解离机制。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号