首页> 外文会议>ASMS Conference on Mass Spectrometry and Allied Topics >A New Dual Cell Linear Ion Trap Configuration for Improved Quadrupole Ion Trap Performance
【24h】

A New Dual Cell Linear Ion Trap Configuration for Improved Quadrupole Ion Trap Performance

机译:一种新的双电池线性离子阱配置,用于改进的四极离子阱性能

获取原文

摘要

A dual cell linear trap configuration enables improved performance over a single cell configuration. A higher pressure region can be utilized for improved efficiency for trapping injected ions, CID, and ion isolation. Trapping efficiencies of > 95percent, CID efficiencies of >80percent, ion isolation resolutions of >7600 with an efficiency of 68percent are achieved. A low pressure region can be utilized to improve mass analysis resolution and/or scan speed. Peak widths of .45 amu can be obtained while scanning at >33,000 amu/sec. Transfer of ions from one cell to another takes < 5 ms and is >90percent efficient for all ions in a 2000 m/z mass range. Further optimization of transfer conditions and other possible advantages of both the high and low pressure regions will continue to be studied in the future.
机译:双单元线性陷阱配置可以通过单个单元配置提高性能。较高的压力区域可用于提高用于捕获注射离子,CID和离子隔离的效率。实现了> 7600效率> 7600的85%,离子隔离分辨率> 7600的诱发效率。可以利用低压区域来改善质量分析分辨率和/或扫描速度。在扫描> 33,000 AMU / SEC时,可以获得.45 AMU的峰宽度。将离子从一个电池转移到另一个电池中需要<5ms,并且在2000m / z质量范围内的所有离子的效率为9。将来将继续研究高压区和低压区域的转移条件和其他可能优势的进一步优化。