A Parallel Three Column System and Direct Pooling of Samples in one Minute: Especially Suitable for in Vitro Screens

机译:一个平行的三列系统,在一分钟内直接汇集样品:特别适用于体外筛选

获取原文
获取外文期刊封面目录资料

摘要

Possibility to pool two samples directly in the injection needle and analyse in 1.6 minutes with good precision (CV 8.7 percent) and low carry over (less than 0.8 percent). The wash cycle of the autosampler and the dead volume of the gradient pump set the limit for the analysis time. About 600 samples can be analysed over night on a single instrument. Hydrophilic compounds like salts and ion suppression agents like MES and HEPES can easily be switched to waste (avoid going to MS). More hydrophobic agents, like PEG, eluting later and thus more difficult to separate from the compounds analysed. The separation can be improved with a longer gradient. With a 3.2 minutes time window it was possible to separate PEG from the compounds analysed.
机译:可能在注射针中直接汇集两个样品,并在1.6分钟内分析良好的精度(CV 8.7%),低携带(小于0.8%)。自动进样器的洗涤循环和梯度泵的死体积设定了分析时间的限制。在单个仪器上可以在夜间分析大约600个样本。盐和离子抑制剂如MES和HEPES等亲水化合物可以很容易地切换到废物(避免进入MS)。更疏水剂,如PEG,以后洗脱,从而更难以分离出分析的化合物。可以通过更长的梯度改善分离。 3.2分钟的时间窗口可以将PEG与分析的化合物分离。