首页> 外文会议>ASMS Conference on Mass Spectrometry and Allied Topics >Maximizing Performance of Sub-2 um Packed Nanobore LC Columns: Minimizing Extra-Column Variance
【24h】

Maximizing Performance of Sub-2 um Packed Nanobore LC Columns: Minimizing Extra-Column Variance

机译:最大化Sub-2 UM Packed Nanobore LC列的性能:最小化额外列方差

获取原文

摘要

All of the post-column, high voltage connection devices tested yielded stable spray throughout the gradient. Minimizing or eliminating post-column volume is a key consideration in the nanobore LC flow path. Maintaining a post-column flow path having an ID less than or equal to the column ID is critical. A larger ID than that of the column is to be avoided. The PicoClear Conductive Union maintained optimal peak shape with a 75 (mu)m ID column packed with sub-2 (mu)m particles. The increase in peak area will be explored for quantitative peptide analysis in determining LOD, LOQ, and linearity.
机译:所有的柱子后,高压连接装置都在整个梯度中产生稳定的喷雾。最小化或消除列后体积是纳米LC流动路径中的关键考虑因素。保持具有id小于或等于列ID的列的列的列流路径至关重要。要避免比列的ID更大。 PicoClear导电联合将最佳峰值保持着75(mU)M ID柱,其填充为亚2(mU)M颗粒。在确定LOD,LOQ和线性度时,将探索峰面积的增加进行定量肽分析。