Review of the Dose and Image Quality Characteristics of 3 FFDM Systems in Clinical Practice in a Screening Programme

机译:在筛选计划中临床实践中3个FFDM系统的剂量和图像质量特征

获取原文

摘要

The purpose of this study was to review the dose and image characteristics of three different FFDM systems; Sectra MDM, GE SenoDS, Lorad Selenia. The dose and image quality characteristics were assessed in terms of both physics and clinical performance. A phantom study was carried out to look at the mean glandular dose delivered to breasts of varying thickness. The dose was also investigated by carrying out a clinical dose survey of a random sample of women on each of the three units. The CDMAM 3.4 was imaged on each unit with varying thicknesses of PMMA and the results correlated with the dose results at each equivalent thickness. The CNR of each unit at varying breast thicknesses was also calculated.
机译:本研究的目的是审查三种不同FFDM系统的剂量和图像特征; Sectra MDM,Ge Senods,Lorad Selenia。在物理学和临床表现方面评估剂量和图像质量特征。进行了幻影研究,以查看递送到不同厚度乳房的平均腺体剂量。通过在三个单位中的每一个上进行随机样品的临床剂量调查,还研究了剂量。 CDMAM 3.4在具有不同厚度的PMMA的每个单元上成像,并且与剂量相关的结果在每个等效厚度相关。还计算了每个单元的CNR以不同的乳房厚度。